SUCTION
There are three major suction system
-
AIR
-
CENTRAL
-
MOTOR
Tokyo Giken is the professional central suction manufacture.
What is central suction?
Several suctions of units can be operated by one powerful suction motor
Benefit
・Operating cost
・Maintenance cost
The feature of Tokyo Giken’ s central suction
Return piping system
Remaining sewerage inside pipes was the biggest concern of central suction system.
All
sewerages will be removed from pipes by installing return piping system.
SUPER BIO FILTER
Exhaust air from suction contains numerous contaminants.
Super bio filter will capture the
contaminants of 0.3 nm or
more at the efficiency of over 99.97%.
Super
bio filter is necessary to maintain clinic environment safe and hygiene.
Without Super bio filter
With Super bio filter
PRODUCT RECOMMENDATION
Recommended suction model
Depends on your clinic size
(How many dental units are
installed?)
unit | suction | |
---|---|---|
6 or more. *Central suction system can be installed on large facility such as hospital or university. |
MAINTENANCE
-
Case 1Suction fails to activate
- Connect firmly or change connector.
-
Case 2Suction activation is slow
【Belt is slipping】- Replace the belt.
-
Case 3Abnormal noise
【Belt is slipping】- Replace the belt.
【 Suction leakage from the separator cover】- Change the separator cover and packing. -
Case 4Suction Pressure is low
【Separator wire-mesh is clogged】- clean wire-mesh
【 The drain valve of the separator is disconnected】- Connect the drain valve
【 Suction leakage from the separator cover or packing】- Change the separator cover and packing
*These are only a part of major causes and troubleshooting.
If the
problem still remains, please contact your distributor
Do not attempt decomposition to
repair by yourself.